Vicker Hardness Tester
An official distributor of Future-Tech in Vietnam
VICKER HARDNESS TESTER
Vicker Hardness Tester FV series is a digital and semi-automatic type. Simple operation and high accuracy.
Features of FV series
FV series is a digital Vickers hardness tester, manufactured by Future-Tech – Japan.
Standout features of Vicker Hardness Tester :
- Advance touch panel operation and display
- Brinell testing
- Fracture toughness testing, Kc
- Clear view and easy measuring
- A variety of selectable loads
- X-Y Coordinate display
- 4 object lenses (Optional)
- 2 indenters for HV/HK Measurement (Optional)
- USB port output
- Equipped with the cylindrical/spherical offset function and the specimen thickness calculation function
- Hardened layer depth calculation function
Technical Information of Vicker Hardness Tester FV series
Loads and Measument system of Machine
TEST FORCE TYPE A
N: 9.807 19.61 29.42 49.03 98.07 196.1 294.2 490.3
kgf: 1 2 3 5 10 20 30 50
TEST FORCE TYPE B
N: 2.942 4.903 9.807 29.42 49.03 98.07 196.1 294.2
kgf: 0.3 0.5 1 3 5 10 20 30
DWELL TIME
FV-110e & FV110: 5 ~ 40 sec
FV-310e & FV-310 & FV-810e & FV-810: 5 ~ 99 sec
TURRET MECHANISM
FV-110e & FV-310e & FV-810e:
Manual
FV-110 & FV-310 & FV-810:
Automatic
DIAMOND INDENTER (Standard)
Vickers Indenter (HV)
DIAMOND INDENTER (Option)
FV-810 & FV-810e: Knoop Indenter (HK) / Two Indenters (HV + HK) are available on the special Dual Indenter Turret simultaneously
OBJECT LENS (Standard)
Type A: X10 / Type B: X10 & X20
OBJECT LENS (Option)
FV-110 & FV-110e & FV-310 & FV-310e: Max. 3 Lenses
FV-810 & FV-810e: Max. 4 Lenses
EYEPIECE
X10
MEASURING MICROSCOPE
Type
FV-110 & FV-110e: Mechanical
FV-310 & FV-310e & FV-810 & FV-810e: Electronic
Max. measurement length
800μm (In case of X 100) /
400μm (In case of X 200)
Graduation
FV-110 & FV-110e: Micrometer: 0.5μm
FV-310 & FV-310e & FV-810 & FV-810e: Digital: 0.1μm (In case of X100 & X200) / 0.01 μm (In case of over X400)
Min. measurement unit
FV-110 & FV-110e: Eye judgment: 0.1μm
FV-310 & FV-310e & FV-810 & FV-810e: Digital: 0.1μm (In case of X100 & X200) / 0.01 μm (In case of over X400)
AVAILABLE TEST PARAMETER
FV-110 & FV-110e & FV-310 &FV-310e: HV / HK
FV-810 & FV-810e: HV / HK / HBS / HBW / Kc
BRINELL HARDNESS TEST
FV-810 & FV-810e: HBS & HBW available Option : Ball indenter & Light load
FRACTURE TOUGHNESS TEST (Kc)
FV-810 & FV-810e: Conform to JIS R1607 / IF Method
HARDNESS CONVERSION
FV-810 &FV-810e: Conform to SAE (J-417b) & ASTM (E-140)
DATA MEMORY
FV-810 & FV-810e: Max. 999 Data
Main Specification of FV series
LOADING MECHANISM
Automatic Load and Release Mothod
LOAD APPLYING SPEED
Initial loading speed : 120 μm / sec Actual loading speed : 60 μm / sec (Automatic speed changing)
MAX. HEIGHT OF SPECIMEN
210 mm
MAX. DEPTH OF SPECIMEN
165 mm
LIGHT SOURCE
LED
OPTICAL FUNCTION
Aperture diaphram/ Field apperture / Color filter : Replaceable
ACCURACY
Conform to JIS B-7725, ASTM E-92 and ISO 6507-2
PHOTOGRAPHIC DEVICE
Capable to mount at anytime on top of the machine (Camera/ Attachment : Option)
DIMENSIONS
(W) 215 x (D) 517 x (H) 690 mm
WEIGHT
Approx. 55 kg (Including Standard Accessories)
POWER SUPPLY
Japan : Single phase AC100V 50/ 60Hz Overseas : Single phase AC100 ~ 240V 50/ 60Hz (Should be specified before shipment)
Display and Output Data
OPERATION PANEL SHEET SW for Type 110 / 310 & TOUCH PANEL SW for Type 810
FV-110e & FV-110
START/ LIGHT UP-DOWN/ TURRET ROTATION (for Auto-Turret Type)
FV-310e & FV-310
START/ RESET / DWELL TIME/ LIGHT UP-DOWN / TURRET ROTATION (for Auto-Turret Type)
FV-810e & FV-810
START/ RESET/ DWELL TIME / LIGHT UP-DOWN/ TEST LOAD / CONVERSION SCALE/ TURRET ROTATION (for Auto-Turret Type) / CLEAR
DATA DISPLAY LED for Type 110 / 310 & LCD for Type 810
FV-110e & FV-110
POWER / LOADING (Pilot Lamp)
FV-310e & FV-310
D1/ D2 / HV-HK/ HARDNESS VALUE / LOADING / OK-NG CRITERIA
FV-810e & FV-810
‘D1/D2 /HV-HK / HARDNESS VALUE / CONVERSION DATA / OK-NG CRITERIA/ START (Blinking for Loading)/ TEST TIMES/etc.
DATA OUTPUT & PRINT OUT
FV-310e & FV-310
RS232C (HV-HK/ HARDNESS VALUE/ OK-NG) USB (D1/ D2 / HV-HK / HARDNESS VALUE/ TEST LOAD/ OK-NG)
FV-810e & FV-810
‘RS232C / USB (D1/D2/HV-HK/ HARDNESS VALUE / TEST LOAD / OK-NG/CONVERSION SCALE / CONVERSION DATA / STATISTIC DATA / etc
OK / NG CRITERIA
Limit setting and HI/ OK/ LO Display
SELF DIAGNOSIS
Display troubles with motors and switches