Load-cell Automatic Vickers Hardness Tester
An official distributor of Future-Tech in Vietnam
LOAD-CELL AUTOMATIC VICKERS HARDNESS TESTER FLV-AR Series
Load-cell Automatic Vickers Hardness Tester is a fully-automatic microhardness
Features of Load-cell Automatic Vickers Hardness Tester
FLV-AS series is a fully automatic Vickers hardness tester series manufactured by Future-Tech – Japan
Outstanding features of the machine
- Load-cell type loading mechanism.
- Load applying mechanism by elevating the head.
- High-resolution camera.
- Sample surface inclination / slope correction function.
- Sample image saving function.
- Large motorized XY stage.
- Various measuring patterns.
- A function that traces the edge of a sample as a standard feature.
- Professional software of image processing.
- Various output data formats and statistical processing.
FT-ZERO Software
Distance & Angle Measurement Function
Only click between two points can measure the distance and click three points can measure the angle easily.
Edge tracing of image function
The original position set is very simple by this function to recognize a shape of sample automatically.
Pattern Setting
Easily set the complicated pattern by clicking a graph. Test load and measurement lens can set it at a Line of each.
Report Graph Chart
Measurement data can save it in CSV form and transfer to an original report.
Technical Information of Load-cell Automatic Vickers Hardness Tester
General specification of Load-cell Automatic Vickers Hardness Tester
FOCUSING METHODS
Automatic Operation (by integrated Auto-Reading Software) / Manual Operation (by motorized Positioning System)
TURRET ROTATION
Automatic-Motorized (by integrated Auto-Reading Software)
MEASUREMENT OF INDENTATION
Automatic Reading / Manual Reading (by integrated Auto-Reading Software)
LOAD APPLYING SPEED
Initial Descending Speed (to 0.15 mm level on specimen surface): Approx. 180 μm/s Effective Loading Speed: Approx. 30 μm/s for under 1kgf Test Load and Approx. 80 μm/s for over 1kgf Test Load
DWELL TIME
5~99sec
INDENTER
For Vickers Test (HV): 1pc (For Knoop Test (HK) and Brinell Test (HBW): Tungsten Ball φ1mm, φ2.5mm (OPTION))
OBJECTIVE LENS
Infinity Corrected Long Working Distance (LWD) Lens: 2pcs X10 & X50
Infinity Corrected LWD Lenses / Max. 5pcs : X1.25, X2.5, X5, X20, X40, X100 (OPTION)
Camera
Automatic measurement by Built-in Type 1.3Mpixel (Made in Germany)
System Control
- Software
FT-ZERO (Windows 10 / 64bit) / Display Size 23”
- Output signal
USB (RS232C Converter: built-in)
- Dimensions
FLV-10AR:
Manual 110 x 110 mm
FLV-10ARS-F:
Auto 150 x 150 mm
- X-Y Stay Movement
FLV-10AR:
Manual (X)50 x (Y)50mm
FLV-10ARS-F:
Auto (X)180 x (Y)110mm
with joystick controller
ACCURACY
For Vickers Test: Conforming to JIS B-7725, JIS B-7734, ASTM E-384 and ISO/ DIN6507-2
SAFETY DEVICE
Protection Cover on Turret for Indenter & Each Objective Lens
SUPPLY
1P AC100~115V / AC240V (50/60Hz): Selectable by SW
Load-cell Automatic Vickers Hardness Tester FLV-10AR and FLV-10ARS-F
Test Load – Vickers
HV 0.01 0.02 0.025 0.05 0.1 0.2 0.3 0.5 1 2 2.5 3 4 5 10 kgf
Test Load – Knoop
HK 0.01 0.02 0.025 0.05 0.1 0.2 0.3 0.5 1 2 kgf
Test Load – Brinell
HBW 1/1 1/2.5 1/5 1/10 1/30 2.5/6.25 mm/kgf
Optional Loading Setting
Possible to set max 64 kinds of test load in a range of 0.01~10kgf by min 0.01kgf.
LOADING SYSTEM
0.01 ~ 10kgf Test load: Automatic Direct Loading System with Load-cell closed-loop feedback
LOAD-CELL TYPE
0.01 ~ 10kgf: by Double Load-cell (Apply Test Load by 10gf)
MAX. HEIGHT OF SPECIMEN
FLV-10AR:
110mm From Manual Stage Top
FLV-10ARS-F:
110mm From Auto Stage Top
MAX. DEPTH OF SPECIMEN
180mm
LIGHT SOURCE
- Bulb
LED Illumination
- Filter
Manual Adapting / Removing
DIMENSIONS
H720 × W415 × D650 mm
WEIGHT POWER
Approx 90kg
Load-cell Automatic Vickers Hardness Tester FLV-50AR and FLV-50ARS-F
Test Load – Vickers
HV 0.01 0.1 0.2 0.25 0.3 0.5 1 2 2.5 3 4 5 10 20 30 50 kgf
Test Load – Knoop
HK 0.05 0.1 0.2 0.3 0.5 1 2 kgf
Test Load – Brinell
HBW 1/1 1/2.5 1/5 1/10 1/30 2.5/6.25 2.5/15.6 2.5/31.25 mm/kgf
Optional Loading Setting
Possible to set max 64 kinds of test load in a range of 0.05~50kgf by min 0.01kgf.
LOADING SYSTEM
0.05~50kgf Test Load: Automatic Direct Loading System with Load-Cell closed-loop feedback.
LOAD-CELL TYPE
0.05~50kgf: by Double Load Cell (Apply Test Load by 10gf)
MAX. HEIGHT OF SPECIMEN
FLV-50AR:
210mm From Manual Stage Top
FLV-50ARS-F:
210mm From Auto Stage Top
MAX. DEPTH OF SPECIMEN
180mm
LIGHT SOURCE
- Bulb
LED Illumination
- Filter
Manual Adapting / Removing
DIMENSIONS
H820 × W415 × D650 mm
WEIGHT POWER
Approx 95kg